IEC 62660-2 130℃ Temperature Cycling Test Chamber Constant  Battery

IEC 62660-2 130℃ Temperature Cycling Test Chamber Constant Battery

Product Details:

Place of Origin: China
Brand Name: LIB
Certification: CE,ROHS
Model Number: TK10-220

Payment & Shipping Terms:

Minimum Order Quantity: 1 Set
Packaging Details: Wooden Cases Standard Export Package
Delivery Time: 5-14 Days
Payment Terms: L/C, D/A, D/P, T/T, Western Union, MoneyGram,Paypal
Supply Ability: 210Sets/quarter
Get Best Price Contact Now

Detail Information

Power Supply: 220V 50Hz / 380V 50Hz Humidity Range: 10%~98%
Temperature Range: -70℃ ~+150 ℃ Heating Rate: 10℃/min (15℃/min)
Name: Temperature Cycling Test Chamber Specimen: Battery
High Light:

130C Temperature Cycling Test Chamber

,

IEC 62660-2 Temperature Cycling Test Chamber

,

130C thermal cycling chamber

Product Description

Description 

IEC 62660-2 130℃ Battery Temperature Cycling Test Chamber are suited for testing battery reliability under extreme temperature changes.

Battery need to be able to withstand drastic changes in temperature.

The thermal shock and temperature cycling tests called out in the IEC, UL, and SAE standards

 

Technical Specification

Name IEC 62660-2 130℃ Battery Temperature Cycling Test Chamber
Model TK10-220
Internal Dimension (mm) 500*600*750
Overall Dimension (mm) 960*1150*1860
Temperature Range (℃) -70℃ ~+150 ℃
Cooling Rate 10℃/min (15℃/min)
Heating rate 10℃/min (15℃/min)
Temperature Fluctuations ± 0.5 ℃
Temperature Deviation ± 2.0 ℃
Interior Volume 220 liters
Controller Programmable color LCD touch screen controller, Multi-language interface, Ethernet , USB
Cooling System Mechanical compression refrigeration system
Exterior Material Steel Plate with protective coating
Interior Material SUS304 stainless steel
Power Supply 380V
Maximum Noise 65 dBA

 

Thermal Test

Battery Temperature Cycling Test Chamber,This test is performed to characterize cell responses to high-temperature environment.

a) Adjust the SOC of the cell to 100% for BEV application, and to 80% for HEV application.

 

b) The cell, stabilized at room temperature, shall be placed in a gravity or circulating air convection oven. The oven temperature shall be raised at a rate of 5K/min to a temperature of 130℃±2K. The cell shall remain at this temperature for 30 min before the test is discontinued.

 

IEC 62660-2 130℃ Temperature Cycling Test Chamber Constant  Battery 0

Note:

If necessary, to prevent deformation, the cell may be maintained during the test in a manner that does not violate the test purpose. The manner to prevent deformation should be representative of cells inside battery systems and battery packs.

 

IEC 62660-2 130℃ Battery Temperature Cycling Test Chamber

 

Test Standard

1. IEC 62660-2:The oven temperature shall be raised at a rate of 5K/min to a temperature of 130℃±2K.

2. IEC 62133 - Temp cycling: 75 to 20 to -20°C (30 min. transitions)

3. UL 2054 - Temp cycling: 70 to -40 °C (30 min. transitions)

4. UL 1642 - Temp cycling: 70 to 20 to -40 °C (30 min. transitions)

5. SAE J2464 - Thermal shock test: 70 to -40°C (15 min. transitions)

IEC 62660-2 130℃ Temperature Cycling Test Chamber Constant  Battery 1

 

The LIB Battery Temperature Cycling Test Chamber meet IEC 62660-2 standard for test battery

 

Contact us for spare parts!

Want to Know more details about this product
I am interested in IEC 62660-2 130℃ Temperature Cycling Test Chamber Constant Battery could you send me more details such as type, size, quantity, material, etc.
Thanks!
Waiting for your reply.